Figure: 25 TAC §289.229(h)(4)(B)(i)

TABLE IV.
HALF-VALUE LAYER FOR SELECTED kVp
  X-ray tube
voltage
(kilovolt peak)
Minimum HVL
(mm of aluminum)
Minimum HVL
(mm of aluminum)
Designed
operating
range
Measured
operating
potential
X-ray systems (except dental)
manufactured before
June 10, 2006
X-ray systems (except dental)
manufactured on or after
June 10, 2006
Below 51 30 0.3 0.3
40 0.4 0.4
50 0.5 0.5
51 to 70 51 1.2 1.3
60 1.3 1.5
70 1.5 1.8
Above 70 71 2.1 2.5
80 2.3 2.9
90 2.5 3.2
100 2.7 3.6
110 3.0 3.9
120 3.2 4.3
130 3.5 4.7
140 3.8 5.0
150 4.1 5.4